Abstract

An automatic microgravimetric screening system based on piezoelectric detection and the use of acidic stannous chloride as reductant was developed for the fast detection and determination of total mercury in water. Reduced mercury is detected as an amalgam by using a gold-coated piezoelectric crystal, the sensor subsequently being regenerated by passing it through a peroxydisulfate solution. The gold-coated piezoelectric crystal is a highly efficient retention unit for the main soluble mercury species (inorganic, complexed, and organometallic) previously reduced to elemental mercury and is free of interferences from other metal ions. This detector exhibits good sensitivity: it allows the determination of mercury at sub-parts-per-billion concentration levels (0.30-1.00 microg/L). The precision, expressed as relative standard deviation, was +/- 2.7% (n = 11; P = 0.05) at 0.5 microg/L total mercury. The proposed method was successfully used as a rapid screening method for mercury monitoring in natural waters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.