Abstract

Silicon (Si) pin diodes can be used for neutron dosimetry by observing the change in forward bias voltage caused by neutron induced displacement damage in the diode junction. Pin diode energy response depends on Si displacement damage KERMA (K(Si)). It is hypothesised that tissue-equivalent (TE) neutron dose could be expressed as a linear combination of K(Si) and foil activation terms. Monte Carlo simulations (MCNP) of parallel monoenergetic neutron beams incident on a cylindrical TE phantom were used to calculate TE dose, K(Si) and Au, Cu and Mn foil activations along the central axis of the phantom. For spectra with neutron energies <100 keV, it is possible to estimate the TE kerma based on silicon damage kerma and Cu or Mn foil measurements. More accurate estimates are possible for spectra where the maximum neutron energy does not exceed 30 keV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.