Abstract

Measurement of the potential of a current-carrying electrode is subject to error when the Luggin capillary of the reference electrode is situated in the path of the current through the electrolyte. This error can be avoided by siting the reference electrode behind a pin-hole through the main electrode, provided that no current flows through the hole. The method also permits the use of a reference electrode of moderate dimensions inserted directly into the cell electrolyte.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.