Abstract
The differential transformer is an important component in the front-end electronics of high-precision capacitive position sensing circuits, which are widely employed in space inertial sensors and electrostatic accelerometers. The position sensing offset, one of the space inertial sensors' most critical error sources in the performance range, is dominated by the differential transformer asymmetry and requires a high-precision evaluation. This paper proposes a method to assess differential transformers' asymmetry and realize a prototype circuit to test a transformer sample. The results show that the asymmetry measurement precision can achieve 0.6 ppm for the transformer with an asymmetry level of about -278.2 ppm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.