Abstract

A method was developed for determining cis-dichlorodiammineplatinum (II) (CDDP) in plasma and urine by HPLC with direct UV detection. After one ml of plasma or urine was pretreated with Dowex 1-x8, an aliquot (100 μ1) of the treated plasma or urine was subjected to HPLC using a column packed with Hitachi Gel #3013-N. A mobile phase was prepared with 10 mM sodium chloride in water/methanol (85 : 15) and the detection wavelength was 210 nm. There was a linear relationship between peak height and the concentration of CDDP in plasma and urine in a range from 1 μM to 500 μM. The detection limit was 0.5 μM of CDDP in plasma and urine. CDDP in the plasma and urine of a patient receiving CDDP treatment can be analyzed by this method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.