Abstract

Limitations of the most commonly used dose calculation methodologies as apply to medium and high-energy x rays are discussed. A method is presented whereby the dose to a point in medium is divided into several components with expressions given for various components. The depth of transient-charged particle equilibrium is chosen as the reference point. Equations are derived for calculating and relating the dose parameters including the percent depth dose, the tissue-reference ratio, and the scatter-reference ratio. Particular emphasis is placed on the accuracy and adequacy of the technique in the region of buildup. The applicability of the present method to megavoltage photons is established for a number of x-ray beams in the range 4-45 MeV maximum energy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.