This paper discusses the development and implementation of a method based on NIR spectroscopy for the in-situ determination of the ammonia mass fraction of ammonia/water mixtures in an absorber test bench. The calibration model was established using a static measuring system. A cell was designed and constructed to prepare and measure samples at the ammonia mass fractions (0.332–0.482), pressures (3.4–4.6), bar and temperatures (25.0–35.5) °C typical in absorption refrigeration systems. A quadratic model for absorbance at 1041nm was established and validated. The root-mean-square deviation (RMSD) of the results was 2.1%. To implement NIR spectroscopy in the absorber test bench, a new flow cell was designed. The calibration model was transferred and used in the conditions of the absorber test bench. In these experimental conditions, the model was statistically validated using density measurements as a reference method for measuring the ammonia mass fraction. The root-mean-square deviation between the ammonia mass fractions obtained using the two methods was 1.1%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.