Abstract
A method by using strain-gauged transducer to adjust the alignment of the materials testing machine is introduced in this paper. In order to improve the accuracy of the material test, the alignment of the material testing machine should be adjusted before tests. A transducer with three levels of strain gauges is applied before the specimens are tested. By calculating the bending stress according to the equation in ASTME1012,an adjustment suggestion to reduce the angular misalignment and concentric misalignment is given by the calculation results. Through adjusting the direction and position of the grips until the value of the strain gauges meet the requirement, the misalignment of the material testing machine will be removed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IOP Conference Series: Materials Science and Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.