Abstract

The emergence of steerable flexible instruments has widened the uptake of minimally invasive surgical techniques. In sinus surgery, such flexible instruments could enable the access to difficult-to-reach anatomical areas. However, design-oriented metrics, essential for the development of steerable flexible instruments for maxillary sinus surgery, are still lacking. This paper proposes a method to process measurements and provides the instrument designer with essential information to develop adapted flexible instruments for limited access surgery. This method was applied to maxillary sinus surgery and showed that an instrument with a diameter smaller than 2.4 mm can be used on more than 72.5% of the subjects' set. Based on the statistical analysis and provided that this flexible instrument can bend up to [Formula: see text] it is estimated that all areas within the maxillary sinus could be reached through a regular antrostomy without resorting to extra incision or tissue removal in 94.9% of the population set. The presented method was partially validated by conducting cadaver experiments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.