Abstract

The ultrasonic cardiac output monitor is a non-invasive, continuous-wave Doppler monitor that can be used to measure cardiac output. Its accuracy and precision in peri-operative and critical care has been assessed with varied results. We conducted a meta-analysis to calculate a pooled mean bias, precision and percentage error for this device. Six studies were suitable for inclusion into a pooled parametric analysis consisting of 320 measurements. The mean weighted bias was -0.39 l.min(-1) (95% CI -0.25 to -0.53 l.min(-1) ), precision 1.27 l.min(-1) and percentage error 42.7% (95% CI 38.5-46.9%). The ultrasonic cardiac output monitor achieved similar agreement with bolus thermodilution to that of other minimally invasive methods of peri-operative cardiac output monitoring, and may have a useful role in cardiac output monitoring.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.