Abstract
We present a MEMS sensor for continuous glucose monitoring for diabetes management. The device consists of a microcantilever, which is driven by remote magnetic field and situated in a microchamber separated from the sensing environment by a semi-permeable membrane. As glucose concentration varies, viscosity changes induced by glucose/copolymer binding in a poly(acrylamide-ran-3-acrylamidophenylboronic acid) (PAA-ran-PAAPBA) copolymer solution produce a measurable change in cantilever vibration. The device has been used to measure physiologically relevant glucose concentrations from 0 to 324 mg/dL. The response time of the sensor to glucose concentration changes was 3 minutes and can be further improved with optimized device designs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.