Abstract

In this paper, a MEMS resonant vacuum gauge (MRVG) based on the squeezed film damping is proposed and its theoretical measurement model is established. The sensitive structure and electrodes of the MRVG were fabricated on SOI wafers and were wafer level packaged including intake hole. An resonance excitation and capacitive detection circuit consists of capacitive voltage converter (C/V) module, PLL and modulation and demodulation module was designed. Experimental tests were carried out in a standard vacuum calibration chamber, and the measurement results verified the theoretical model. The indication accuracy of the MRVG is less than ±20% in the vacuum range of 9×10−5 Pa ∼ 1 Pa. Moreover, the output voltage is still not saturated at the highest vacuum level (9×10−5 Pa), which lays the foundation for subsequent measurement of higher vacuum levels.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.