Abstract

This paper reports a real-time motion-sensing module, which is realized by incorporating multiple MEMS accelerometers into a standard Foley catheter, for assisting diagnosis and treatment of stressed urinary incontinence. The accelerometers measure the orientations of the catheter at multiple points, so the shape of the urethra and movement of the bladder neck can be tracked in real time. An algorithm for extracting tilting, position and shape information from 3-axis MEMS accelerometers has been developed. The model of measurement errors for both static and dynamic testing is also established. The experimental results indicate that the module tracks the movement of the Foley catheter successfully in a real-time environment and the absolute error for static measurement is no more than 1.1° within the operation range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.