Abstract

Numerous experimental and computational methods have been developed to estimate tissue elasticity. The existing testing techniques are generally classified into in vitro, invasive in vivo and non-invasive in vivo. For each experimental method, a computational scheme is accordingly proposed to calculate mechanical properties of soft biological tissues. Harmonic motion imaging (HMI) is a new technique that performs radio frequency (RF) signal tracking to estimate the localized oscillatory motion resulting from a radiation force produced by focused ultrasound. A mechanical model and computational scheme based on the superposition principle are developed in this paper to estimate the Young's modulus of a tissue mimicking phantom and bovine liver in vitro tissue from the harmonic displacement measured by HMI. The simulation results are verified by two groups of measurement data, and good agreement is shown in each comparison. Furthermore, an inverse function is observed to correlate the elastic modulus of uniform phantoms with amplitude of displacement measured in HMI. The computational scheme is also implemented to estimate 3D elastic modulus of bovine liver in vitro.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.