Abstract
In dynamic temperature measurement, thermocouple sensors are widely used, and their dynamic characteristics directly affect the accuracy of the test results. So before applying thermocouple sensors to dynamic temperature measurement, their dynamic characteristics should be obtained, and their dynamic performance parameters should be analyzed. The time constant is the most important dynamic parameter, which reflects the response speed of a thermocouple sensor. Therefore, it is necessary to measure the time constant. The time constant is closely related to the heat transfer mode, so the heat transfer environment of the time constant measurement system should be similar to the application environment of thermocouple sensor. When using the thermocouple to measure the temperature in various kilns, the heat transfer is mainly through the radiative mode, and existing equipment, such as constant temperature water/oil tank, shock wave tube and other devices, cannot be used to measure the time constant to reflect scenarios in actual measurement applications. Therefore, this paper proposed a new method to measure the time constant of the thermocouple by improved high temperature furnace. In the system, the high temperature furnace was used to generate the stable temperature field, and the fast feed device was used to insert the thermocouple into the high temperature furnace and generates the ramp temperature excitation. The temperature can reach 1500 °C in the furnace, and the temperature error in uniform temperature field is ±1 °C. Finally, the time constant of a K-type thermocouple was measured, and the uncertainty of the measurement result was analyzed.
Highlights
Thermocouple sensor is widely used in dynamic temperature measurement in harsh environment because of its simple structure, low cost, and strong robustness
When the temperature of the uniform temperature zone is within 300–700 ◦ C, the temperature profile from the furnace mouth to uniform temperature zone is linear
As can be seen from the table, if the measurement error of the time constant is less than 5%, the rise time of the ramp excitation signal should be less than 10% of the time constant
Summary
Key Laboratory of Instrumentation Science & Dynamic Measurement (North University of China), Ministry of Education, Taiyuan 030051, Shanxi, China.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.