Abstract
A measurement system for evaluating ultrasound fields is presented. The system utilizes a PVDF hydrophone and a mechanical scanning device. For continuous wave (CW) ultrasound, a high sensitivity is achieved by using a narrow band technique. The high sensitivity is demonstrated in a recording of the beam profile from an air transducer. Recordings from fetal Doppler equipment shows that, due to interference between the multiple transmitters, a high spatial resolution is needed in the field measurements in order to obtain correct values for peak intensities.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.