Abstract

We have designed a phantom to evaluate mean glandular dose (MGD) as part of the regulatory dosimetry control for mammographic equipment. The phantom is constituted by TLD-100 thermoluminescent dosemeters (TLDs) inserted within semicircular plates of acrylic. Different groups of TLDs are used to determine entrance surface air kerma and half-value layer (HVL). Calibration of both tasks has been performed using a Senographe 2000D system and an ionization chamber. The phantom has been tested in five clinical systems. The HVL and MGD obtained by this method agree, on average, within 3%, with those from standard procedures based on the use of ionization chambers. The phantom MGD measurements have a combined uncertainty better than 10% (k = 1).