Abstract
Abstract A computerized adiabatic calorimeter for heat capacity measurements in the temperature range 80–400 K has been constructed. The sample cell of the calorimeter, which is about 50 cm3 in internal volume, is equipped with a platinum resistance thermometer and surrounded by an adiabatic shield and a guard shield. Two sets of 6-junction chromel-copel thermocouples are mounted between the cell and the shields to indicate the temperature differences between them. The adiabatic conditions of the cell are automatically controlled by two sets of temperature controller. The reliability of the calorimeter was verified through heat capacity measurements on the standard reference material α-Al2O3. The results agreed well with those of the National Bureau of Standards (NBS): within 0.2% throughout the whole temperature region. The heat capacities of high-purity graphite and polystyrene were precisely measured in the interval 260–370 K by using the above-mentioned calorimeter. The results were tabulated and plo...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.