Abstract

This paper introduces a piezoresistive microcantilever with a low spring constant. The microcantilever was fabricated with titanium (Ti) as the piezoresistor, a low spring constant polyimide (PI) layer, and a thin silicon oxide (SiO2) layer as the top and bottom passive layers, respectively. Excellent mechanical performances with the spring constant of 0.02128 N/m and the deflection sensitivity of 1.03 × 10−7 nm−1 were obtained. The output voltage fluctuation of a Wheatstone bridge, which consists of four piezoresistive microcantilevers, is less than 3 μV@3 V in a phosphate buffered saline (PBS) environment. A microcantilever aptasensor was then developed through functionalizing the microcantilevers with a ricin aptamer probe, and detections on ricin with concentrations of 10, 20, 50 and 100 ng/mL were successfully realized. A good specificity was also confirmed by using bovine serum albumin (BSA) as a blank control. The experiment results show that the Ti and PI-based microcantilever has great prospects for ultrasensitive biochemical molecule detections with high reliability and specificity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.