Abstract
We have designed a new low-pass Fresnel zone plate (LPFZP) to extend soft x-ray absorption spectroscopy (XAS) to the lithium K absorption edge in a scanning transmission x-ray microscope (STXM). The performance of the LPFZP was evaluated in the STXM beamline at the UVSOR-III Synchrotron (Okazaki, Japan); the contribution of the higher-order harmonics is successfully suppressed to 0.1% of the fundamental energy, and a spatial resolution of 72 nm and an energy resolution (E/∆E) above 1000 are achieved as expected. XAS spectra of lithium are measured successfully in an electrode of a lithium-ion battery.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.