Abstract

SummaryThe UK influenza pandemic plan predicts up to 750 000 additional deaths with hospitals prioritising patients against inadequate resources. We investigated three prototype low‐cost, gas‐efficient, pneumatic ventilators in a test lung model at different compliance and rate settings. Mean (SD) oxygen consumption was 0.913 (0.198) and 1.119 (0.267) l.min−1 at tidal volumes of 500 ml and 700 ml respectively. Values of F I o 2 increased marginally as lung compliance reduced, reflecting the increased ventilator workload and consequent increased enrichment of breathing gas by waste oxygen from the pneumatic mechanism. We also demonstrated that a stable nitric oxide concentration could be delivered by this design following volumetric principles. It is possible to make a gas‐efficient ventilator costing less than £200 from industrial components for use where oxygen is available at 2‐4 bar, with no pressurised air or electrical requirements. Such a device could be mass‐produced for crises characterised by an overwhelming demand for mechanical ventilation and a limited oxygen supply.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.