Abstract

This paper presents a novel approach of using a miniaturized potentiostat (M-P) chip (LMP91000) to perform full range cyclic voltammetry (CV) measurements for the detection of biomarkers. The LMP91000 evaluation board was reconfigured to perform three-electrode CV measurements in order to achieve electrochemical cortisol immunosensing. The microelectrodes for cortisol estimation were fabricated by immobilizing monoclonal anti-cortisol antibody (Anti-M-Cab) onto self-assembled monolayer (SAM) modified Au microelectrodes. The results obtained using the M-P were compared to those obtained using a conventional potentiostat. The M-P was successful in measuring cortisol levels in the range of pM. The outcomes of the studies suggest that M-P can effectively perform biochemical measurements on three electrode systems, enabling the development of miniature systems for point-of-care (POC) diagnosis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.