Abstract
BackgroundNeonatal jaundice, defined as an excess of bilirubin in blood at birth or shortly thereafter, has a disproportionately high impact in low-resource settings. Three quarters of deaths from jaundice occur in sub-Saharan Africa and south Asia. Jaundice is easily treated with blue light phototherapy; however, accurate, low-cost bilirubin monitoring is essential to guide effective phototherapy treatment in low-resource settings. To address this need, we developed BiliDx, a low-cost, portable reader that quantifies bilirubin concentration in a few drops of blood applied to a lateral flow card. BiliDx measures absorbance at three wavelengths: blue (bilirubin), amber (haemoglobin interference), and red (background) and can perform measurements in less than 5 mins at the point of care, without the need for a costly centrifuge or extensive training. Here, we compare the accuracy of BiliDx against a standard laboratory bilirubinometer and a transcutaneous bilirubinometer. MethodsThis study took place in two central hospitals in Malawi. We included samples from neonates being treated in neonatal wards and deemed to be at risk for jaundice or who were undergoing phototherapy. Guardians provided written informed consent. We measured bilirubin using three devices: BiliDx, a bilirubinometer (Reichert UNISTAT), and a transcutaneous bilirubinometer (Drager JM-105), with a total of 475 measurements per device. Training (149 measurements) and validation (326 measurements) sets were created by ordering the first measurement per patient from lowest to highest UNISTAT bilirubin; every third measurement was selected for the training set. The training set was used to fit a model that correlated bilirubin concentration to absorbances at three wavelengths. The validation set was used to make an unbiased evaluation of the accuracy of the model. FindingsWe enrolled 456 neonates between August 7 and December 12, 2019; 247 of 326 (75%) of BiliDx measurements in the validation set were within 2 mg/dL of UNISTAT measurements, compared with 44 of 281 (6%) of quantitative transcutaneous measurements. No adverse events occurred. InterpretationBiliDx can provide accurate bilirubin monitoring. BiliDx is more accurate than transcutaneous bilirubinometry, a technology often endorsed for its non-invasive, portable nature. Limitations include that only ten nurses used BiliDx and only central hospitals were included. Testing of the BiliDx device is ongoing; we aim to collect enough measurements >20 mg/dL to assess the accuracy of BiliDx over the intended bilirubin concentration range of 0–35 mg/dL. Future studies will evaluate usability among a larger group of nurses and hospital settings. FundingUSAID Saving Lives at Birth Award No. 7200AA18FA0016.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.