Abstract

A mass spectrometer has been designed to distinguish ions of different mass but identical mass to charge ratio. Separation of mass from charge is based upon secondary particle emission when each ion impacts the detector. The yield of secondary particles is a function of kinetic energy rather than energy-per-charge. Combined with an electrostatic energy gate and time-of-flight, the mass can be determined from kinetic energy, energy-per-charge, and mass-per-charge. The design is of moderate precision but simpler than available alternatives – particles need not be extracted from the vacuum chamber for sizing nor the detector kept at cryogenic temperatures. The prototype is demonstrated on an aluminum ablation plume and an ionic liquid ion source.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.