Abstract

We report a longitudinal Zeeman slower based on ring-shaped permanent magnetic dipoles used for the strontium optical lattice clock. The Zeeman slower is composed of 40 permanent magnets with the same outer diameter but different inner diameters. The maximum variation of the axial field from its target values is less than 2%. In most parts of the Zeeman slower, the intensity variations of the field in radial spatial distribution are less than 0.1 mT. With this Zeeman slower, the strontium atoms are slowed down to 95 m/s, and approximately 2% of the total atoms are slowed down to less than 50 m/s.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.