Abstract

ABSTRACTThis electrode system for electrocardiography and impedance pneumography is an extension of the one used on Project Mercury, Detailed description is given of the manufacturing procedure for the silver‐silver chloride electrodes, for the preparation of a non‐irritating electrode paste and for the application of the electrodes. Tests were made on 21 subjects up to 90 hours. Only nine electrodes failed during the entire test during which subjects engaged in active sports, bathed, etc. Irritation of the skin was minimal.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.