Abstract
Design and performance are given of a multiple traversal low temperature absorption cell that operates routinely at a path length of 230 m. The system is comprised of an f/30-aperture White cell surrounded by an aluminum Dewar. The cell has been used in the temperature range 300-77 K and may be pressurized to 3 atm. Vacuum feedthrough adjustments are employed in order to correct for defocusing of the mirrors during the cooling process.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.