Abstract
We present an experimental setup developed to measure locally the fluctuations of themagnetization of physical systems such as spin and superspin glasses. It is based onmicronic and submicronic Hall probes. We present the noise reduction at ambienttemperature owing to the use of the spinning current technique. Finally, we show why, withsuch probes, the noise measured on a macroscopic sample probes only a microscopicvolume of the sample.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Statistical Mechanics: Theory and Experiment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.