Abstract

A linear magnetic field scan driver was developed to provide a rapidly scanning magnetic field for use in electron paramagnetic resonance (EPR) spectroscopy. The driver consists of two parts: a digitally synthesized ramp waveform generator and a power amplifier to drive the magnetic field coils. Additionally, the driver provides a trigger signal to a data collection digitizer that is synchronized to the ramp waveform. The driver can also drive an arbitrary current waveform supplied from an external source. The waveform generator is computer controlled through a serial data interface. Additional functions are controlled by the user from the driver front panel. The frequency and amplitude of the waveform are each separately controlled with 12-bit resolution (one part in 4,096). Several versions of the driver have been built with different frequency and amplitude ranges. Frequencies range from 500 to 20,000 Hz. Field sweep amplitudes range up to 80 G(pp). This article also gives a brief description of the field coils that are driven by the driver.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.