Abstract

Abstract Cardiovascular, orthopedic, and interventional radiology procedures using fluoroscopy require healthcare professionals to wear heavy lead garments for radiation protection, which can lead to musculoskeletal injuries, discomfort, and fatigue. A mobile lead garment frame, MobiLead, was designed to mitigate these issues by moving the lower garment load off body to a structural frame and redistributing the load of the upper garment from the shoulders to the hips through a torso frame. The frame was designed to be compact with a minimal footprint for maneuverability in operating rooms with limited space, while still giving the surgeon adequate mobility for various emergency procedures. Preliminary analysis of device efficacy was conducted using electromyography (EMG) of load-bearing muscles during use and qualitative surgeon user feedback surveys. Nonparametric permutation tests showed that the MobiLead device mean activation was significantly lower than the standard lead garment (p<0.001). Surgeon feedback and use testing showed that the device fits under a sterile gown and is maneuverable in an operating room (OR).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.