Abstract

A high-performance liquid chromatography tandem mass spectrometry cubed (HPLC/MS3) method was developed and validated to quantify lamotrigine in human plasma with carbamazepine as an internal standard. The HPLC/MS/MS system is composed of a Shimadzu UFLC XR high-performance liquid chromatograph coupled with a hybrid linear ion trap triple quadrupole mass spectrometer. Following simple protein precipitation with methanol, the separation of lamotrigine and carbamazepine was performed on an Agilent Poroshell 120 SB-C18 column (4.6 × 50 mm, 2.7 μm) using gradient elution with 0.1% formic acid in water (solvent I) and 0.1% formic acid in methanol (solvent II) at a flow rate of 0.8 mL min-1. The total run time for each sample was 5 min. The method was validated for accuracy, precision, linearity, lower limit of quantification (LLOQ), selectivity, and other parameters. The LC/MS3 method was linear in the concentration range of 0.50-50.0 μg mL-1 (R2 ≥ 0.995). The LLOQ was 0.5 μg mL-1, requiring only 30 μL of human plasma. Intra- and inter-day accuracies were <6.17% and precisions were <11.4% at all concentrations. The absolute recoveries (%) and matrix effect (%) for lamotrigine in human plasma were between 83.8 and 90.7. The developed and validated LC-MS3 assay was successfully applied to monitor the lamotrigine levels in human plasma after the administration of lamotrigine.