Abstract

A rapid and highly sensitive method for the determination of acamprosate (ACM), in human plasma using ESI-LC-MS/MS (electrospray ionization liquid chromatography tandem mass spectrometry) in negative ionization polarity in multiple reactions monitoring (MRM) mode was developed and validated. The procedure involves a simple protein precipitation step. Chromatographic separation was carried out on a Hypersil BDS C(18) column (150 mm × 4.6 mm, 5 µm) with an isocratic mobile phase and a total run time of 2.5 min. The standard calibration curves were linear within the range of 7.04-702.20 ng/mL for ACM (r ≥ 0.990). This study briefly describes the role of ion source design on matrix effects. ACM shows matrix effects in z-spray ionization source design, whereas it has no matrix effects in orthogonal spray ion source design. This method was successfully applied to a pharmacokinetic study after oral administration of acamprosate 333 mg tablet in Indian healthy male volunteers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.