Abstract

This paper shows that the resolving power of the retina of the human eye when any type of cataract is present can be determined by means of a He-Ne laser retinometer (λ=633 nm). The results of a measurement of the retinal visual acuity of more than two hundred patients with various types of cataract before extraction of the crystalline lens and after replacement and the determination of the corresponding correction coefficient, reaching 0.7-0.9, make it possible to use the laser retinometer for pre-operative diagnosis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.