Abstract

Twenty-five patients with peripheral circulatory disorders were examined by laser Doppler and strain-gauge techniques to compare these methods of measuring distal systolic blood-pressure. The correlation coefficients for simultaneous measurement of toe and ankle systolic blood-pressures were high, 0.98 and 0.99 respectively. The correlation coefficient for successive measurements of toe pressure was lower, 0.83, probably because of a variation in blood-pressure with time. The laser Doppler method seems to be more sensitive than the strain-gauge method in the low-pressure range. The laser Doppler probe is easy to attach to most skin surfaces and the laser Doppler technique may be, therefore, an alternative and a complement to the strain-gauge method when the strain-gauge is difficult to use on damaged or ulcerous toes and feet.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.