Abstract
SPEED-Mk.II, the second KAWAI-type high P–T apparatus for in situ X-ray observation that was installed at the bending magnet beam line at SPring-8, is described. The guide block system was designed so that the change of the relative dimension of the cubic compression space with press load can be minimized by repeated adjustments. The hydraulic system was designed so as to enable smooth compression and decompression. These precise controls should be advantageous for high-pressure generation, especially when sintered diamond (SD) anvils are used. An oscillation system was equipped for the first time in a large volume press, making it possible to obtain high-quality diffraction patterns even when the number of sample grains is limited. The use of the oscillation system also reduces errors in pressure determination that may be caused by insufficient averaging of diffraction angles over grains in a limited diffraction volume, because the oscillating grains should sweep through the 2θ range that is allowed by the finite widths of the optical slits.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.