Abstract

The gas system for the Outer Tracker of the HERA-B experiment at DESY produces the desired counting gas mixture Ar/CF 4/CO 2 65:30:5 and circulates it through the detector at a flow rate of 20 m 3/ h , i.e. ∼1 vol/ h . It controls flows and regulates pressures in all 26 OTR half-superlayers, purifies the gas upon return from the detector, and automatically performs a quantitative analysis of main and trace (O 2, N 2, H 2O) gas components for the common input and the outputs of all half-superlayers. The first running experience and the strategies employed during system construction to avoid any detector aging possibly induced by the gas system are discussed. The large system with major gas purification stations was constructed using only non-outgassing, “clean” materials and devices, such as stainless steel, PEEK, baked Viton, and metal bellows pumps. An epoxy glue was used extensively as a non-outgassing sealing material in applications with up to 100 bar pressure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.