Abstract

The construction and performance of a large drift chamber are described, which has a sensitive area of 5 m 2 and an average chamber thickness of 0.55 × 10 −3 radiation lengths. Simultaneous sense wire- and high voltage read-out gives a good time correlation with ajitter of 2.0 ns fwhm. Measurements with a collimated radioactive α-sources resulted in a time resolution of 4.3 ns fwhm corresponding to a space resolution of 0.20 mm fwhm at the sense wire. The space resolution at the high voltage plane is about 25 mm fwhm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.