Abstract
An automatic sample changer has been developed for use in Ge(Li) gamma ray spectroscopy. The sample changer features remote storage to prevent cross-talk and to reduce background. It has a capacity for 200 samples. It also features automatic setting of the sample to detector distance to optimise the count rate performance. An additional feature is the rotation and vibration of samples during counting to assure that powdered samples are compacted and improve the precision and reproducibility of the counting geometry. Three systems have been equipped with the changer and these are connected to an in-house computer on a first-come-first-served basis via a high speed data interface (153,6Kb).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.