Abstract

The lamp described comprises a 1000 W incandescent filament source, long- and short-wave pass filters, and a non-imaging reflector system. The output of 24 W in the 620-720 nm band is delivered through a 53 mm diameter aperture with a maximum divergence half-angle of 60�. Refracting components may be fitted to modify the output angular or intensity distribution. This lamp has been used to irradiate several different types of malignant tumour in human patients, following intravenous injection of the photoactive drug haematoporphyrin derivative, causing selective necrosis of the malignant tissue. The clinical results are regarded as encouraging.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.