AbstractThe accurate measurement of trace element concentrations in natural sulphides by laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) has been limited by the lack of matrix-matched calibration standards. The synthesis of a standard, IMER-1, by incorporating four minor and 34 trace elements into a chalcogenide glass matrix Ge28Sb12S60is reported here. Chemical analysis by electron probe microanalysis (EPMA), LA-ICP-MS, solution ICP-MS, and inductively coupled plasma-optical emission spectroscopy (ICP-OES) confirmed the excellent homogeneity of major elements (1-σ relative standard deviation (RSD) <1% for S, Sb and Ge) and acceptable homogeneity of most trace elements (1-σ RSD <10%). The standard was validated by analysing trace-elements concentrations in three geological pyrite specimens using IMER-1 as the calibration standard and comparing the results to previously reported values also determined by LA-ICP-MS but using a different calibration standard. STDGL2b-2. The results suggest that IMER-1 may be an appropriate calibration standard for LA-ICP-MS analysis of trace elements in natural sulphides.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.