This paper reports a flexible, inductive wireless sensor that can be combined with a variety of hydrogel materials for biomedical applications. The variable inductor is formed by folding coplanar dual spiral coil with 5-10 mm size that are microfabricated using 50-mum-thick copper-clad polyimide film so that when folded, the two coils are aligned to each other and the mutual inductance depends on the gap between the aligned coils. A hydrogel element is sandwiched by the folded substrate to modulate the gap, or inductance of the device as it swells/deswells depending on the target parameter. The response of a variable inductor to the displacement of the coils is measured to be 0.40 nH/mum. A sensitivity of 113 ppm/mum in wireless frequency measurement is obtained using the passive resonant device that combines the variable inductor with a fixed capacitor. The resonant devices are coupled with a commercial hydrogel wound dressing as well as pH-sensitive poly(vinyl alcohol)-poly(acrylic acid) hydrogel. Wireless monitoring of the swelling of the wound dressing is experimentally demonstrated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.