Abstract

Hybrid small-diameter tubes were fabricated by wrapping decellularized aortic intima-media sheets around a tubular stainless steel mandrel with diameter 4 mm, and then by coating with electrospun segmented polyurethane. The synthetic coat was deposited uniformly to a thickness of about 0.5-3.5 μm depending on the duration of electrospinning. Resistance to luminal pressure, burst strength, and stiffness increased with the thickness of the electrospun coat, suggesting that the synthetic fabric reinforces the reconstructed acellular aortic intima-media. Human umbilical vein endothelial cells seeded on the inner surface acquired flagstone morphology, while normal human dermal fibroblasts seeded on the outer surface proliferated well and partly migrated into deeper layers. Collectively, the data suggest that reinforcing decellularized aortic intima-media with electrospun fibers generates a small-diameter hybrid blood vessel with good biocompatibility and suitable mechanical properties. © 2019 Wiley Periodicals, Inc. J Biomed Mater Res Part A: 107A: 1064-1070, 2019.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.