Abstract
1. A calorimeter for the continuous measurement of heat production and heat loss in the human subject, for at least 24 h, is described. The calorimeter operated on the heat-sink principle for direct calorimetry and an open-circuit system for indirect calorimetry. 2. Sensible heat loss was measured using a water-cooled heat exchanger, and the temperature of water entering the heat exchanger was controlled to maintain a mean temperature gradient of zero across the chamber walls. 3. Evaporative heat loss was determined from ingoing and outgoing wet-and-dry bulb temperatures and air flow-rates. 4. Problems associated with the calculation of evapoative heat loss and the estimation of the volume of incoming air in open-circuit systems are considered. 5. The calibration, limits of accuracy, sources of error and experiments with subjects are discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.