ABSTRACT A low-cost and computer-controlled graphitization system connected to an elemental analyzer (EA) has been designed and built at the NTUAMS Lab. This semiautomatic system equips 6-unit reactors for the graphitization of CO2 with H2 on the iron catalyst. The entire procedure takes about 7 hours for iron conditioning, sample combustion and loading, and graphitization. The system can produce good-quality graphite for samples containing 0.5–1.6 mg carbon mass, with the pressure yield of graphitization ranging from 57.7% to 87.1%. The average values of OXI and OXII agree well with the consensus value, but the result of ANU sucrose was observed to be slightly higher than the reported one. The background samples of anthracite over ten months yielded an average of 0.38±0.10 pMC (n=21) corresponding to a 14C age of 45 kyr BP. Intercomparison samples L and M of FIRI exhibit that the measured 14C ages are almost identical to the consensus values and have a small spread in these values. The system has been carrying out graphitization for total organic carbon (TOC) of peat samples, and providing a more efficient and convenient way for AMS 14C dating.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.