Abstract
We developed a hollow-fibre column system specifically adapted to prepare washed platelet concentrates (WPCs). This study was performed to evaluate the efficacy of the hollow-fibre column system for preparing WPCs. First, the percentages of platelet (PLT) recovery and remaining plasma proteins were calculated by determining the PLT count, volume and plasma protein levels in both the prewash and postwash. Secondly, washed PLTs and unwashed control PLTs were stored for 5 days, and the changes during this 5-day storage of in vitro PLT characteristics were determined. The hollow-fibre column system effectively removed >98% of plasma in platelet concentrates (PCs), and the PLT recovery was 97% on an average. The CD62P-expression level on washed PLTs immediately after washing was approximately twofold higher than that on prewashed PLTs as well as on PLTs washed via manual methods or cell washing devices. Until day 5 during storage, PLT aggregability, hypotonic shock response and swirling scores of washed PLTs were not significantly different from those of the control PCs. Our novel hollow-fibre column system proved valuable in preparing washed PLTs with <2% of residual plasma proteins and high recovery of PLTs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.