Abstract
A sensor of $m$ -toluidine polymer film coated platinum electrode was fabricated by the electropolymerization using cyclic voltammetry technique for the detection of mercury ions (Hg2+) in aqueous solution. This paper was carried out using the simple potentiometric method and confirmed by a cyclic voltammetry technique. The effects of the polymer film thickness and pH of Hg2+ solutions on the response of the sensor were studied. Moreover, the stability, sensitivity, and selectivity of the $m$ -toluidine sensor were investigated. The optimum thickness of the polymer film was obtained after ten cyclic voltammetric runs. This film has a Nernstian response slope of 29.19 mV/decade with a detection limit of $3.54\times 10^{\mathrm {\mathbf {-5}}}$ M at 293 K by the simple potentiometric method. In addition, it has a sensitivity of $4\times 10^{\mathrm {\mathbf {-7}}}$ AM $^{\mathrm {\mathbf {-1}}}$ with a detection limit of $1.33\times 10^{\mathrm {\mathbf {-7}}}$ M by the cyclic voltammetry method. Moreover, the sensor is specific to Hg2+ ions in the presence of other ions, such as Na+, K+, Mg2+, Ca2+, Co2+, Ni2+, Zn2+, and Pb2+. Furthermore, the most stable response of the sensor to Hg2+ ions in the solution of pH ranged from 4 to 6.4 for a lifetime of about eleven weeks. Moreover, the sensor was applied for detection of four natural samples: 1) tap water; 2) underground water; 3) first distillate water; and 4) another sample wasted with Hg2+ ions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.