Abstract
We developed an electrochemical enzymatic biosensor coated with red blood cell membrane (RBCM) for highly selective glucose measurement. The RBCM containing its membrane proteins (glucose transporter-1) was extracted from a human red blood cell (RBC), and then placed on an established enzymatic glucose sensor by the vesicle fusion method. In the newly fabricated glucose sensor, the RBCM plays the role in the diffusion barrier of preventing the penetration of interfering molecules, while in contrast making glucose molecules permeable. We employed scanning electron microscopy and atomic force microscopy to analyze the RBCM diffusion barriers. The performance of our glucose sensors with diffusion barriers was tested regarding selectivity to glucose against other interfering molecules, such as ascorbic acid, uric acid, and galactose. Remarkably, employing the RBCM significantly improved the selectivity and accuracy of the glucose measurement, even under human serum, compared to the uncoated sensors. This result implies that the RBCM serves well as a highly permselective layer to glucose molecules, and a diffusion barrier to other molecules.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.