Abstract

This study aimed to develop a streamlined method for evaluating the dilution ratio of drug dose–response plates created by automated liquid handlers in the early stages of drug discovery. The quantitative techniques commonly used for this purpose have restrictions due to their limited linear dynamic range and inaccuracies in assessing serial dilution performance. To address this challenge, we describe a method based on acoustic ejection mass spectrometry (AEMS). The method involves using standard compounds and an internal standard to evaluate each dilution point in quality control (QC) plates. The samples are transferred to a chromatography-free tandem mass spectrometry system through an acoustic source, enabling the analysis of one sample per three seconds from a microtiter plate. This approach provides precise, accurate, label-free, and rapid data acquisition to support high-throughput screening efforts.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.