Abstract
The high intrinsic brightness of the circulating electron beam at SURF-II is used as the entrance slit for a two-meter normal incidence monochromator. A typical beam size for the electron beam is 100 μ high by 2 mm wide yielding an observed resolution of 0.4 Å with a 200 μm exit slit and a 2400 lines/mm grating. The instrument accepts a beam with a 65 mrad horizontal divergence and a 10 mrad vertical divergence. A plane pre-mirror used near normal incidence reflects the incoming radiation onto the 2 m grating; this combination provides a horizontal exit beam, and enables the experiment to be located three meters from the orbit tangent point. With magnesium fluoride coated aluminum optics a flux of 2×10″ photons/s·Å at 1200 Å is observed with a 10 mA circulating current. A flux of 5×10 10 photons/sdÅ at 600 Å is observed with an osmium coated grating and a 10 mA circulating current. Sample spectra of the angle-resolved photoelectron spectrum of CO are presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.