Abstract

Accurate measurement of the height of individual pulses in a signal that is a.c.-coupled is dependent upon correcting a changing baseline to a constant value. A new fast baseline restorer, using internally triggered gated feedback with gain, has been designed and a version optimized for Gaussian-shaped pulses. Performance is dependent on the magnitudes of both the forward and the feedback gains. The design presented can achieve a 0.5% accuracy in pulse height for a minimum pulse width of 2.5 μs or for 5 μs overlapping pulses as long as the overlap is less than 50%. Typical applications for the baseline restorer include blood cell volume measurement systems, scintillation counters and radiation spectroscopy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.